Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
Mechanical Data
Wall-/Pipe-mount versions
IP66/NEMA4X
Dimensions 192mm high x 230mm wide x 94mm deep
(7.56 in. high x 9.06 in. wide x 3.7 in. deep)
Weight 1kg (2.2 lb)
Panel-mount versions
IP66/NEMA4X (front only)
Dimensions 96mm x 96mm x 162mm deep
(3.78 in. x 3.78 in. x 6.38 in. deep)
Weight 0.6kg (1.32 lb)
Cable Entry Types
Standard
5 or 7 x M20 cable glands
North American
7 x knockouts suitable for 1/2 in. Hubble gland
Power Supply
Voltage requirements
85 to 265V AC 50/60 Hz
24V AC or 12 to 30V DC (optional)
Power consumption
<10VA
Insulation
Mains to earth (line to ground) 2kV RMS
IM/AX4PH
Issue 9
Environmental Data
Operating temperature limits
-20 to 65ºC (-4 to 149ºF)
Storage temperature limits
-25 to 75ºC (-13 to 167ºF)
Operating humidity limits
Up to 95%RH non condensing
EMC
Emissions and immunity
Meets requirements of:
EN61326 (for an industrial environment)
EN50081-2
EN50082-2
Hazardous area approvals
CENELEC ATEX IIG EEx n IIC T4
FM non-incendive Class I Div. 2 Groups A to D
CSA non-incendive Class I Div. 2 Groups A to D
Safety
General safety
EN61010-1
Overvoltage Class II on inputs and outputs
Pollution category 2
Languages
Languages configurable:
English
French
German
Italian
Spanish
9 SPECIFICATION
Pending
Pending
Pending
SS/AX4PH Issue 5
67

Advertisement