4.1.2 Set Up User Defined Buffers (pH Only)

Buffer Type
set to User
(see Section 4.1.1)
15. 0
Deg.C
0. 0 0
pH
A: Enter point 1
15. 0
Deg.C
3. 9 9
pH
A: Enter point 1
35. 0
Deg.C
4. 0 2
pH
A: Enter point 5
15. 0
Deg.C
3. 9 9
pH
B: Enter point 1
35. 0
Deg.C
4. 0 2
pH
B: Enter point 5
Set Auto Buffers
IM/AX4PH
Issue 9
Solution A: Enter point 1 (to 5)
and
Deg.C
Adjust
keys, adjust the temperature reading (in 5º increments) to the first of the
temperatures on the pH/temperature curve.
and
are shown alternately on the center display line. Using the
pH
Adjust
keys, adjust the pH reading (in 0.01pH increments) to the pH reading that
corresponds to the temperature reading entered above.
Notes.
1)
For accurate calibration, it is important to repeat the above for buffer solution A
at all 5 points along the pH/temperature curve.
2)
The displayed temperature value increases automatically by 5ºC from the value
set for the previous point. The setting may be increased but not decreased.
Solution B: Enter point 1 (to 5)
Solution B set up is identical to solution A set up.
Note. For accurate calibration, it is important to repeat the above for buffer solution B at
all 5 points along the pH/temperature curve.
The analyzer calculates the pH/temperature relationship from the data entered.
See Section 4.1.3.
Sensor Cal. A
SECURITY CODE
Alter Sec. Code
CONFIG. DISPLAY
Alter Sec. Code
are shown alternately on the upper display line. Using the
not set to zero (Section 5.8) - see section 5.1, page 28.
set to zero (Section 5.8) - see section 5.2, page 29.
4 SETUP
and
and
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax436Ax460Ax466Ax468

Table of Contents