ABB AX416 User Manual page 64

Single and dual input analyzers for ph/redox orp
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
Option Board
Fitted and Analog
Features Enabled
-----
Cal. Output 3
-----
Cal. Output 4
Option Board Fitted,
Additional Features
Disabled
Analog
-----
Pb Dp
Option Board
Option Board
Not Fitted
0000
Alter Fact. Code
62
Calibrate Output 3
Output 3 (and Output 4) calibration is applicable only if the Option board is fitted
and analog features enabled - see below.
Output 3 calibration is identical to Output 2 calibration.
Calibrate Output 4
Note. Output 4 calibration is identical to Output 3 calibration.
Configure Option Board
Notes.
This frame is displayed only if an option board is fitted.
The software detects if an option board is fitted but cannot detect the additional
features available.
If an option board is fitted, the correct selection must be made below to enable use
of the available features. If an incorrect selection is made, the software menus and
frames associated with that option are displayed in the Operating and Configuration
pages but the features do not work.
Use the
Analog
Pb Dp
Analog + Pb Dp
Alter Factory Code
Set the factory settings access code to a value between 0000 and 19999.
Return to main menu.
FACTORY SETTINGS
Monitoring pH
Redox (ORP)
Operating Page - see section 2.3, page 6.
Dual Redox
pH/Redox (ORP)
and
keys to enable the features for the type of option(s) fitted:
-
Analog features enabled (comprising two additional
analog outputs, two additional alarm relays, clock and
logbook facility).
-
PROFIBUS-DP digital communications features
-
enabled.
Both analog and PROFIBUS-DP features enabled.
7 CALIBRATION
IM/AX4PH
Issue 9

Advertisement