Physiological Traces - GE Vivid S70 User Manual

Hide thumbs Also See for Vivid S70:
Table of Contents

Advertisement

Prepare the System for Use

Physiological traces

CAUTION
CAUTION
3-14
The physiological module consists of two channels: ECG and
Respiration. The scanned image that is displayed is
synchronized with the ECG and respiration traces. In M-Mode or
Doppler, the traces are synchronized to that particular mode's
sweep.
Use only GE Healthcare accessories.
Conductive parts of electrodes and associated connectors for
applied parts, including neutral electrodes should not contact
other conductive parts, including earth.
Simultaneous use of two or more applied parts will cause
summation of patient leakage currents.
Heart rate may be adversely affected by cardiac arrhythmias or
the operation of cardiac pacemaker pulses.
Figure 3-8. The patient (I/O) connection panel
Vivid S70 / S60
User Manual
BC092760-1EN 01