Prepare the System for Use
Physiological traces
CAUTION
CAUTION
3-14
The physiological module consists of two channels: ECG and
Respiration. The scanned image that is displayed is
synchronized with the ECG and respiration traces. In M-Mode or
Doppler, the traces are synchronized to that particular mode's
sweep.
Use only GE Healthcare accessories.
Conductive parts of electrodes and associated connectors for
applied parts, including neutral electrodes should not contact
other conductive parts, including earth.
Simultaneous use of two or more applied parts will cause
summation of patient leakage currents.
Heart rate may be adversely affected by cardiac arrhythmias or
the operation of cardiac pacemaker pulses.
Figure 3-8. The patient (I/O) connection panel
–
Vivid S70 / S60
User Manual
BC092760-1EN 01