GE Vivid S70 User Manual page 56

Hide thumbs Also See for Vivid S70:
Table of Contents

Advertisement

Safety
Guidance and manufacturer's decleration - electromagnetic immunity.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF and Proximity fields
from RF wireless
communications equipment
IEC 61000-4-3
NOTE: Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
2-24
IEC 60601-1-2 test level
3 Vrms
150 kHz to 80 MHz
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
385 MHz (18 Hz Pulse
Modulation)
450 MHz (FM +/ -5 kHz deviation
1 kHz sine or 18 Hz Pulse
Modulation)
710 MHz (217 Hz PM)
745 MHz (217 Hz PM)
780 MHz (217 Hz PM)
810 MHz (18 Hz PM)
870 MHz (18 Hz PM)
930 MHz (18 Hz PM)
1720 MHz (217 Hz PM)
1845 MHz (217 Hz PM)
1970 MHz (217 Hz PM)
2450 MHz (217 Hz PM)
5240 MHz (217 Hz PM)
5500 MHz (217 Hz PM)
5785 MHz (217 Hz PM)
Compliance level
3 Vrms
150 kHz to 80 MHz
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
27 V/m
28 V/m
9 V/m
9 V/m
9 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
9 V/m
9 V/m
9 V/m
Vivid S70 / S60