Measurements and Analysis
CAUTION
Exiting AFI
8-36
Rejection of non-physiological traces and correct peak
detection (see also 'Inspecting results' on page 8-31) is
particularly important when using the TTP color map, as wrong
peaks will influence the PSD index significantly. Peaks
detected in very early systole and late diastole should be
checked, and traces rejected if they are considered
non-physiological.
Figure 8-19. Bull's Eye and Traces screen
At any time, the tool can be cancelled by pushing the Cancel
button. The plugin will close and any unsaved data will be lost.
When in the Results stage the operator may click the Approve
and Exit button. The behavior depends on the number of views
analyzed:
•
If all three views have been analyzed, the tool will go to the
Bull's eye only layout and prompt the operator whether the
results should be stored. If Yes is selected, measurements
–
Vivid S70 / S60
User Manual
BC092760-1EN 01