GE Vivid S70 User Manual page 646

Hide thumbs Also See for Vivid S70:
Table of Contents

Advertisement

Probes
Chemicals Used for Efficacy Validation
Table 13-4:
Product
Type
Trade Name
Enzymatic
Enzol®
Detergent
(Cidezyme®)
(Soak)
MetriZyme™
Prolystica® 2X
Concentrate
Presoak &
Cleaner
Intermediate-l
Oxivir® Tb
evel
Disinfectant
(wipe)
High-level
Cidex® OPA
Disinfectant
(Soak)
McKessen
OPA/28
13-30
The instructions provided above have been validated to properly
prepare GE Ultrasound probes for re-use. It remains the
responsibility of the processor to ensure that the processing, is
performed as specified in this document. This may require
verification and routine monitoring of the process.
The table below lists the products and intended use (clean, ILD,
HLD) that were validated.
Chemicals used for Efficacy Validation on non-TEE probes
Manufacturer
Advanced
Sterilization
Products®
(J&J)
Metrex™
Steris
Diversey
Advanced
Sterilization
Products (J&J)
McKesson
Minimum
Contact
Active
Time
Ingredient
1-Minute
Proteolytic
Soak
Enzymes
10-Minute
Hydrogen
Exposure
Peroxide
10-Minute
Ortho-
Soak
phthalaldehyde
Vivid S70 / S60
Validated for
probe
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
3Sc-RS,
M5Sc-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
User Manual
BC092760-1EN 01

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s60

Table of Contents