Creation of the formula
1. Select the last measurement.
2. Double click and enter the formula name.
3. Select "=" to create the formula.
Vivid S70 / S60
User Manual
BC092760-1EN 01
Figure 12-13. The Add measurement window
3. MCO and AV ET are measurements that already exist on
the system, check Use copy of and select MCO from the
drop down menu.
4. Select OK to add the MCO measurement.
5. Repeat steps 2 to 4 to add the AV ET measurement.
Figure 12-14. The Measurement menu sheet
The formula for this example is as follows:
My LIMP = (MCO-AV ET)/AV ET
1. In the user-defined folder (e.g."My folder"), select the last
measurement created (e.g. AV ET).
2. Click beneath the last line in the Parameter list in the
Measurement menu sheet to add a new line.
Measurement Package