Measured value output after signal through the "Contact"
or "Pulse" inputs
To start measured value output through the EXT interface
(X41), you can either:
ä
Close the "Contact" input (pin 23 on X41) against 0 V,
for example with a simple switch (make contact);
or
ä
Close the "Pulse" input (pin 22 on X41) against 0 V,
for example by triggering the input with a TTL logic
device (such as SN74LSxx).
Characteristic times for measured value output
Process
Minimum duration of "Contact" signal
Minimum duration of "Pulse" signal
Storage delay after "Contact"
Storage delay after "Pulse"
Measured value output after
Regeneration time
Duration of measured value transfer
187 + (11 • number of blank lines)
t
=
D
baud rate
Time
Triggering the "Contact" and "Pulse" inputs at D-sub
connection EXT (X41)
³
t
7 ms
e
³
t
1.5 µs
e
£
t
5 ms
1
£
t
1 µs
1
£
t
57 ms
2
³
t
0
3
[s]
Signal transit times for measured value output after "Pulse"
or "Contact"
EXT(X41)
Pin 23
Pin 1(0V)
EXT(X41)
Pin 22
Pin 1(0V)