Working with the position display unit

Advertisement

Quick Links

HEIDENHAIN
Working with the position
display unit
Actual value and input display
(7-segment LED,
9 decades and sign)
REF
<
=
HEIDENHAIN
Status display
• Switch display to:
MIN / MAX / DIFF / ACTL / START / PRINT
• Go to parameter list after switch-on
• Page forward in parameter list
Indicator
REF
in.
1 /
2
PRINT
SET
< / = / >
MIN / MAX
DIFF / ACTL
START
• Select datum
• Page backward in
1
2
SET
START
>
MIN
ACTL
Meaning
If the decimal points have stopped blinking:
Reference mark was crossed over — datum points are now stored
in nonvolatile memory.
Blinking: Waiting for operator to press ENT or CL.
Position values displayed in inches.
Datum 1 / Datum 2 currently active.
Blinking: Waiting for ENT for data output.
Blinking: Waiting for operator to confirm entry.
Sorting mode: Measured value less than lower limit / within
tolerances / greater than upper limit.
Measuring series: Minimum / Maximum /
largest difference (MAX–MIN) / current measured value.
Blinking: Waiting for confirmation of value to be displayed.
Measuring series in progress.
Blinking: Waiting for start signal for measuring series.
ND 286
parameter list
PRINT
in.
MAX
DIFF
• Clear entry
• Reset to zero (P80!)
• CL plus MOD:
parameter list
• CL plus 2-digit num-
ber: select parameter
• Clear parameter entry
and show parameter
number
• Confirm entry value
• Set display to value from
P79 (P80!)
Numeric keypad
7
8
9
4
5
6
1
2
3
MOD
.
CL
0
• Algebraic
sign
• Decrease
parameter
value
• Decimal point
• Increase pa-
rameter value

Advertisement

Table of Contents
loading

Summary of Contents for HEIDENHAIN ND 286

  • Page 1 HEIDENHAIN Working with the position ND 286 display unit • Confirm entry value • Set display to value from Actual value and input display • Select datum P79 (P80!) (7-segment LED, • Page backward in 9 decades and sign) Numeric keypad...
  • Page 2: Setting The Datum

    The ND 286 position display unit is designed for use with HEIDENHAIN linear or rotary encoders with square-wave output signals. These encoders have one or more reference marks, which may also be distance-coded . When a reference mark is crossed over, it generates a signal identifying that position as a reference point.
  • Page 3: Distance-To-Go Mode

    Measuring Series The ND 282 display unit can calculate and display one of the following values from a measuring series: Smallest value (MIN), largest value (MAX), difference between largest and smallest value (DIFF), last value measured (ACTL) A new value is captured every 550 µs during a measuring series. To start a measuring series: Press the MOD key repeatedly until the desired indicator starts blinking.
  • Page 4: Data Output

    Output instantaneous values referenced to datum 1 (MIN/MAX/DIFF display values are not output) A connecting cable (to a PC, for example) is available from HEIDENHAIN (Id.-Nr. 206 420 ..); cable length up to 10 m (32.8 ft). Operating parameters for data output: P23, P53 to P57 “...
  • Page 5 D-Sub Connection EXT (25-pin, male) Danger to internal components! Voltage sources for external circuitry must conform to the recommendations in EN 50 178 for low-voltage electrical separation. Connect inductive loads only with a quenching diode parallel to the inductance. Use only shielded cable! Connect the shield to the connector housing.
  • Page 6: Error Messages

    Internal counter overflow Error while crossing over reference marks To clear the error message: Switch off the display unit. Should any of these error codes recur, contact your HEIDENHAIN service agency. Erase the operating parameters. If all decimal points light up, the measured value is too large or too small.
  • Page 7: Operating Parameters

    Operating Parameters The parameters are divided into “user parameters“ and “protected operating parameters," which can only be accessed by entering a code number. User parameters User parameters are operating parameters that you can change without entering the code number: They are designated P00 to P30, P79, P86 Calling user parameters To call user parameters immediately after switch-on: Press the MOD key as long as...
  • Page 8 Parameter Meaning Function / Effect Setting Subdivision of encoder signals Subdivision 4, 2, 1, 0.8, 0.5, 0.4, 0.2, 0.1 Counting 0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - 0 mode 0 - 2 - 4 - 6 - 8 - 0 0 - 5 - 0...
  • Page 9 Fast external reset/preset Fast Set repeated (setting of P53: external REF mode, datum 2 and reset/preset measuring series are not usable No fast preset Parameter Settings for HEIDENHAIN Linear Encoders Model Signal Reference P 43 Display step The following settings period marks...
  • Page 10: Rear Panel

    100 ... 240 V 50 ... 60 Hz Power switch Ground connection Input for HEIDENHAIN linear encoder with square-wave output signals (TTL), connecting cable max. 50 m (164 ft), minimum edge separation 175 ns The X1, X33 and X41 interfaces comply with EN 50 178 recommendations for separation from line power.