HEIDENHAIN ND 281B User Manual page 48

Measured value displays
Table of Contents

Advertisement

Operating parameters for measured value output
Parameter
Baud rate
P50 RS232
Number of additional blank lines for
P51 RS232
measured value output
Display freeze during measured value output
In operating parameter P23, you can specify how the
measured value output signal will affect the display unit.
Display freeze during measured value output
Concurrent display, no display freeze: The
display value is the current measured value
Frozen display: Display is stopped
(frozen) and updated by every
measured value output signal
Frozen/concurrent display: Display is
frozen as long as a measured value output
signal is present
48
Function
DISPL. ACTL.
DISPL. HOLD
DISPL. STOP
Measured value output via PRINT function
In the linear measurement mode, press MOD repeatedly
until the PRINT indicator starts blinking.
Start the measured value output with ENT.
In the angular measurement mode, press MOD (this feature
can be disabled with operating parameter 86).
Duration of measured value transfer
187 + (11 • number of blank lines)
t
=
D
baud rate
Indicator preselection (linear measurement mode)
Operating parameter P86 allows you to define which indicator
P23
is displayed first when MOD is pressed.
[ s]