Switching Inputs and Outputs EXT (X41)
Danger to internal components!
Voltage sources for external circuitry must conform
to the recommendations in EN 50 178 for low-
voltage electrical separation. Connect inductive
loads only with a quenching diode parallel to the
inductance.
Only use shielded cables!
Connect the shield to the connector housing.
Outputs at D-sub connection EXT (X41)
Pin
Function
14
Display value is zero
15
Measured value
16
Measured value
17
Measured value < lower sorting limit (P18)
18
Measured value > upper sorting limit (P19)
19
Error (see "Error Messages")
38
1)
Only in the linear measurement mode
³
trigger limit A1 (P62)
³
trigger limit A2 (P63)
Inputs at D-sub connection EXT (X41)
Pin
Function
1, 10
0 V
2
Reset display to zero, clear error message
3
Set display to the value selected in P79
4
Ignore reference mark signals
5
Start series of measurements
6
Externally select display value for series of
measurements
7
Display MIN value of series of measurements
8
Display MAX value of series of measurements
9
Display difference MAX – MIN
22
Pulse: Transmit measured value
23
Contact: Transmit measured value
25
Enable or disable REF mode
(current REF status is changed)
12, 13, 24
Do not assign
11, 20, 21
Vacant
Special case: Display current measured value ACTL
If you wish to display the current measured value ACTL of a
series of measurements, note for inputs 7, 8 and 9:
Either none or more than one of these inputs must be active.
1)
1)
1)
1)
1)