Siemens SITRANS F Operating Instructions Manual page 146

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Alarms and system messages
9.2 Alarm messages
Siemens
NAMUR
ID
132
135
136
139
140
143
144
147
148
149
151
152
192
193
195
196
198
199
200
201
202
144
Diagnostic
Corr. volumeflow above upper
alarm limit
Corr. volumeflow below lower
alarm limit
Totalizer 1 above upper alarm
limit
Totalizer 1 below lower alarm
limit
Totalizer 2 above upper alarm
limit
Totalizer 2 below lower alarm
limit
Totalizer 3 above upper alarm
limit
Totalizer 3 below lower alarm
limit
Transmitter temp. above upper
alarm limit
Transmitter temp. below lower
alarm limit
Sensor serial number mismatch
Transmitter serial number
mismatch
Dosing time overrun
Dosing quantity overrun
Current output value below
Lower Scaling
Current output value above
Upper Scaling
Frequency output value below
Lower Scaling
Frequency output value above
Upper Scaling
Pulse overflow
Current output value below
Lower Scaling
Current output value above
Upper Scaling
Action
Check process conditions or align limit to normal
operation. Adjust parameter "Upper Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Lower Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Upper Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Lower Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Upper Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Lower Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Upper Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Lower Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Upper Limit Alarm"
Check process conditions or align limit to normal
operation. Adjust parameter "Lower Limit Alarm"
SensorFlash backup is disabled due to mismatch of
serial numbers between sensor front-end and
SensorFlash
SensorFlash backup is disabled due to mismatch of
serial numbers between transmitter and SensorFlash
Check installation. If ok, increase "Duration Time"
Check installation. If ok, decrease "Overrun Value"
Check process conditions or align limit to normal
operation. Adjust channel 2 parameter "Lower Scaling"
Check process conditions or align limit to normal
operation. Adjust channel 2 parameter "Upper Scaling"
Check process conditions or align limit to normal
operation. Adjust channel 2 parameter "Flow Value
Low"
Check process conditions or align limit to normal
operation. Adjust channel 2 parameter ""Flow Value
High"
Pulse output insufficient pulse separation. Increase
"Amount Per Pulse" or reduce "Pulse Width" on
channel 2
Check process conditions or align limit to normal
operation. Adjust channel 3 parameter "Lower Scaling"
Check process conditions or align limit to normal
operation. Adjust channel 3 parameter "Upper Scaling"
Operating Instructions, 03/2012, A5E03361511-01
SITRANS FC430 with HART