Oem Menu; Maximum Temperature - Thermo Scientific Barnstead Pacific RO 3 Operating Instructions Manual

Water purification system
Table of Contents

Advertisement

12.2 OEM Menu

Basic settings and limiting values can be changed in this menu.
To make changes in the OEM menu, the system control must previously be unlocked (see
11.1.8).
Calling the OEM menu:
Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu. Following
this, the prompt "OEM menu Press enter" appears. When this is confirmed with the Enter-key,
the first menu prompt can be worked on. To simplify changing settings, press the UV-key to se-
lect the number in a value which you want to change. Now use the arrow keys to enter a number
from 0 to 9 at the selected position.
A press on the menu-key takes you to the next menu prompt.
The OEM menu prompt display shows:

12.2.1 Maximum temperature:

A single press on the menu-key:
(this is not active in this version)
The maximum temperature which the system can be allowed to reach can be set in this menu.
When this temperature is exceeded, the "max. Temp." fault message is triggered. Settings above
50 °C cause the limiting value to be suppressed and the word „off" appears in the 4
display.
Basic setting:
Setting range:
The display shows:
28
OEM menu
Press Enter !
50 °C
1 - 50 °C
OEM menu
max. Temp.
35 °C