Operator's Manual Simple AFG Application
Trigger In signal (blue, top) that starts a burst of sine waveform (red, bottom)
Reference Clock Input Connector
When the "Clock Source" is set on "External" in the "Device Settings" page, the internal clock
synthesizer uses the signal from "Reference Clock Input" SMA connector to generate the DAC sampling
clock signal.
External Modulation Input Connector
The T3AWG-3K series instrument can accept the modulating signal from an external source, through the
External Modulation Input SMA connectors located on the Rear Panel.
External Modulating Input Description
Connector Type
Input Impedance
Input Voltage Range
Reference Clock Output
Reference Clock Input
Connector type
Input impedance
Input Frequency range 5 MHz to 100 MHz
1 SMA on Rear Panel for all channels
10 kOhm
±0.5 V for all modulations
Value
1 SMA on the Rear Panel
50 Ohm, AC coupled
48