Marker Out Behaviour In Continuous Run Mode - Teledyne T3AWG3K Series Operator's Manual

Hide thumbs Also See for T3AWG3K Series:
Table of Contents

Advertisement

Operator's Manual Simple AFG Application

Marker Out behaviour in Continuous Run Mode

In Continuous mode, the Marker Out generate a pulse with a duty cycle of 50% at the beginning of each
period. The Marker have the same frequency of the carrier waveform until it is below 75 MHz. Over
75MHz the Marker Out frequency is divided by an integer value as described in the following:
Marker out frequency = Carrier frequency / 1 when: Carrier frequency < 75MHz
Marker out frequency = Carrier frequency / 2 when: 75MHz < Carrier frequency <= 150MHz
Marker out frequency = Carrier frequency / 4 when: 150MHz < Carrier frequency <= 300MHz
Note: for the Noise and DC waveforms, the Marker Out is unavailable in continuous mode.
Marker out (blue, top) synchronous with the analog signal (red, bottom)