Operation and Maintenance
October 2018
4.6.4
LCD display
Transmitters ordered with the LCD display will be shipped with the display installed.
Installing the display on an existing Rosemount 3051SMV requires the LCD display kit (part
number 03151-9193-0001 for aluminum housing and 03151-9193-0004 for stainless steel
(SST) housing).
About this task
Use the following procedure and
Procedure
1. If the transmitter is installed in a loop, then secure the loop and disconnect power.
2. Remove the transmitter cover on the feature board side (opposite the field
3. Engage the four-pin connector into the feature board and snap the LCD display into
4. Install the display cover and tighten to ensure metal to metal contact in order to
4.6.5
Flange and drain vent
The Rosemount 3051SMV is attached to the process connection flange by four bolts and
two alignment cap screws.
Procedure
1. Remove the two alignment cap screws.
124
terminals side). Do not remove the instrument covers in explosive environments
when the circuit is live.
place.
meet flameproof/explosion-proof requirements.
Figure 4-13: Optional LCD Display
A. Feature board
B. LCD display
C. Display cover
Figure 4-13
to install the LCD display:
A
B
Rosemount 3051S Multivariable Transmitter
Reference Manual
00809-0100-4803
C