Liquiphant FTL41
5.2

Mounting the device

5.2.1
Required tool
• Open-ended wrench for sensor installation
• Allen key for housing locking screw
5.2.2
Installation
Aligning the tuning fork using the marking
 9
Markings to align the tuning fork
Installing in pipes
• Flow velocity up to 5 m/s with a viscosity of 1 mPa⋅s and density of 1 g/cm
Check for correct functioning in the event of other process medium conditions.
• The flow will not be significantly impeded if the tuning fork is correctly aligned and the
marking is pointing in the direction of flow.
• The marking is visible when installed.
 10
Installation in pipes (take fork position and marking into consideration)