Mounting The Device - Endress+Hauser Liquiphant FTL64 Brief Operating Instructions

Point level switch for liquids in high-temperature applications
Hide thumbs Also See for Liquiphant FTL64:
Table of Contents

Advertisement

Liquiphant FTL64 HART
 8
Examples of support in the event of dynamic load
Marine approval: In the case of pipe extensions or sensors longer than 1 600 mm (63 in),
a support is needed at least every 1 600 mm (63 in).
5.2

Mounting the device

5.2.1
Required tool
• Open-ended wrench for sensor installation
• Allen key for housing locking screw
5.2.2
Aligning the tuning fork using the marking
The tuning fork can be aligned using the marking in such a way that the medium drains off
easily and buildup is avoided.
Markings are identified on the process connection by:
Material specification, thread designation, circle, line or double line
Endress+Hauser
Mounting
A0042356
13