Mounting
5.1.2
Take viscosity into consideration
Viscosity values
• Low viscosity : < 2 000 mPa⋅s
• High viscosity: > 2 000 to 10 000 mPa⋅s
Low viscosity
It is permitted to position the tuning fork within the installation socket.
 3
Installation example for low-viscosity liquids. Unit of measurement mm (in)
High viscosity
NOTICE
Highly viscous liquids may cause switching delays.
Make sure that the liquid can run off the tuning fork easily.
Deburr the socket surface.
The tuning fork must be located outside the installation socket!
 4
Installation example for a highly viscous liquid. Unit of measurement mm (in)
8
> 25 (0.98)
> 40 (1.57)
Liquiphant FTL41
A0033297
A0037348