Safety notes
Table 202 from EN 60601-1-2:2001
Guidance and manufacturer's declaration – electromagnetic immunity
The Leica M525 OH4 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the Leica M525 OH4 surgical microscope should make sure that it is used in
such an environment.
Immunity test
Discharge of static
electricity (ESD) according to contact discharge
IEC 61000-4-2
High-freq. transient
electrical interference /
Burst
according to IEC 61000-4-4
Surges
according to
IEC 61000-4-5
Voltage drops
short interruptions
and voltage variations
IEC 61000-4-11
Magnetic fields at
mains frequency
(50/60Hz) according to
IEC 61000-4-8
Note:
62
IEC 60601 test level
Compliance level
± 6KV
± 6KV
contact discharge
± 8KV air
± 8KV air
± 2KV for
± 2KV for
power lines
power lines
± 1KV for input and
± 1KV for input and
output lines
output lines
± 1KV
± 1KV
normal mode
normal mode
± 2KV
± 2KV
common mode
common mode
<5% UT (>95%
<5% UT (>95%
dip in UT)
dip in UT)
for
1 1
/
cycle
for
2 2
40% UT (60%
40% UT (60%
dip in UT)
dip in UT)
for 5 cycles
for 5 cycles
70% UT (30%
70% UT (30%
dip in UT)
dip in UT)
for 25 cycles
for 25 cycles
<5% UT (>95%
<5% UT (>95%
dip in UT)
dip in UT)
for 5 sec.
for 5 sec.
3 A/m
not applicable
U
is the a.c. mains voltage prior to application of the test level.
T
Leica M525 OH4 / Ref. 10 714 367 / Version -
Electromagnetic environment – guidance
1 1
/
cycle
2 2
Floors should be of wood, concrete or
ceramic tile. If the floor is of synthetic mate-
rial, the relative humidity must be at least
30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Leica M525 OH4 surgical
microscope requires that the instrument
remain functional even after mains power
interruptions, it is recommended that the
Leica M525 OH4 surgical microscope be
provided with an auxiliary power source
such as an uninterruptible power supply
(UPS) or battery back-up.