Use
The red line on the brightness adjustment bar shows the
maximum adjustable brightness for the current working
distance.
The brightness cannot be set to a level beyond the red line
When the working distance is reduced by too little at a set
brightness, the brightness is reduced automatically.
This safety function can be deactivated by clicking the
"Brightcare™" button. A dialog window opens in which
you have to confirm that you want to deactivate the
safety function.
40
When the "Brightcare™" safety function is deactivated, the
color of the "Brightcare™" button changes from green to yellow.
Reactivating the "Brightcare™" safety function:
➩Click the yellow "BrightCare™" button again.
"BrightCare™" is now activated and the button is again lit green.
Leica M525 OH4 / Ref. 10 714 367 / Version -
Warning 7
At a short focal distance, the light source of the
illumination unit may possibly be too bright for the
operating physician and the patient.
➩Start with a low light level and gradually increase the
light intensity until the operating physician has an opti-
mally illuminated image.
The status of the "BrightCare™" safety function can
only be changed permanently in the "User settings"
menu. A change in status during operational
procedures will not be stored when the user settings
are saved with "Save" or "Save as"!