Preparation for operation
➩Check the balancing.
➩Press the "All Brakes" button on the CAN handle.
The optical carrier tilts to the right:
➩Move the C-axis to the left with switch (1) until the optical
carrier is balanced.
The optical carrier tilts to the left:
➩Move the C-axis to the right with switch (1) until the optical
carrier is balanced.
The optical carrier (in the A-position) tilts back:
➩Move the A-axis forwards with switch (2) until the optical
carrier is balanced.
32
1
2
Leica M525 OH4 / Ref. 10 714 367 / Version -
The optical carrier tilts forwards:
➩Move the A-axis back with switch (2) until the optical carrier
is balanced.
3
The optical carrier (in the B-position, e.g. for posterior fossa
operations) tilts back:
➩Move the B-axis forwards with switch (x) until the optical
carrier is balanced.
The optical carrier tilts forwards:
➩Move the B-axis back with switch (x) until the optical carrier
is balanced.
If the microscope cannot be balanced manually, the
weight of the accessories is probably outside the
balanceable weight range. This can only be done for A,
B and C axes by reducing or increasing the accessory
weight to within the permitted range (see page 80).
.