➩The following dialog window opens on the touch panel:
The balancing procedure can be canceled at any time
using "Interrupt Balancing".
➩Balancing is complete when the acoustic signal no longer
sounds and the push-button (2) is no longer flashing.
A dialog window indicates that balancing has been completed.
➩Press the "Close" button or wait until the dialog window is
closed automatically after 5 seconds.
➩Check the balancing.
➩Press the "All Brakes" button on the CAN handle and position
the microscope.
The microscope must remain fixed in any position.
If automatic balancing cannot be successfully completed, the
following dialog window opens:
➩Acknowledge with the "Close" button.
➩Correct the alignment of the optical carrier (B-position).
➩Press the push-button (2).
Autobalancing re-starts.
Leica M525 OH4 / Ref. 10 714 367 / Version -
Preparation for operation
29