Safety notes
Caution 1
Danger of injury due to falling weight disk or cover!
➩When changing the weight disk, make sure that your
feet are not beneath the weight disk or the cover.
Caution 2
Surgical microscope can move without warning!
➩Always lock the footbrake when you are not moving
the system.
Caution 3
Damage to the Leica M525 OH4 surgical microscope
due to uncontrolled tilting!
➩ Hold the CAN handles securely before you press the
"All Brakes" button.
Caution 4
Damage to the Leica M525 surgical microscope during
transportation!
➩Never move the stand in the extended condition.
➩Never roll over cables lying on the floor.
Caution 5
Risk of infection!
➩Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
Caution 6
Hot lamp insert can cause burns!
➩Do not touch the hot lamp insert.
Caution 7
Heating of tissue
If the illumination field diameter is larger than the
visual field and the light intensity is set too high, uncon-
trolled heating of tissue can occur outside the visible
range of the microscope.
➩Do not set the light intensity too high.
Caution 8
Destruction of the zoom motor!
➩Use the manual adjustment of the zoom motor only if
the zoom motor is defective.
Caution 9
Damage of the touch panel
➩Operate the touch panel using your fingers only.
Never use hard, sharp or pointed objects made out of
wood, metal or plastic.
Caution 10
Damage of the touch panel
➩Never clean the touch panel using cleaners that
contain abrasive substances. These substances can
scratch the surface and cause it to be become dull.
60
Leica M525 OH4 / Ref. 10 714 367 / Version -
Caution 11
Danger of skin burns!
The lamp insert is very hot.
➩Before changing the lamp, make sure that the lamp
insert has cooled down.