➩Press the start button (1) for automatic AC/BC balancing on
the swing arm.
During the balancing procedure, the push-button (2) on the con-
trol unit flashes yellow and an acoustic signal sounds.
The B-axis is balanced first, followed by the C-axis.
➩Check the balancing.
➩Press the "All Brakes" button on the CAN handle and position
the microscope.
The microscope must remain fixed in any position.
➩Correct the balancing, if necessary.
Correcting the balancing manually
For manual balancing, the axes can be moved manually using
switches (3), (4) and (5).
1
2
Leica M525 OH4 / Ref. 10 714 367 / Version -
5
4
Make sure that no accessories collide with the
microscope during manual balancing.
Preparation for operation