8. Connect the plug of the fluorescence cube ⑤ in one of the
free connectors ⑥ to power the LED. (Fig. 13)
9. Apply the adhesive marker ⑦ for the fluorescence cube on
the illuminator. (Fig. 14)
10. Close the top door.
11. Close the side cover.
12. Connect the power supply.
13. Start working.
8.1.3
Replacing a fluorescence filter
1. Disconnect the power supply plug from the fluorescence il-
luminator.
2. Open the side cover of the illuminator, by unscrewing the
side screws ①. (Fig. 9)
It could be helpful to remove the observation head.
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slide and
vice versa.
3. Open the top door of the fluorescence illuminator by un-
screwing the four screws ② and release the cover. (Fig. 10)
4. Loosen the front locking screw ③ on the fluorescence cube
slider. (Fig. 11)
5. Disconnect the plug ⑤ related to the cube you want to re-
place. (Fig. 13)
6. Remove the fluorescence cube from the dovetail ④ of the
cube slider.
7. Repeat steps 5. to 9. of paragraph 8.1.2 to install a new flu-
orescence cube.
Page 14
F ig. 13
F
ig. 13
F ig. 14
F
ig. 14