8.1.2
Installing additional fluorescence filter
1. Disconnect the power supply plug from the fluorescence il-
luminator.
2. Open the side cover of the illuminator, by unscrewing the
side screws ①. (Fig. 9)
It could be helpful to remove the observation head.
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slider and
vice versa.
3. Open the top door of the fluorescence illuminator by un-
screwing the four screws ② and release the cover. (Fig. 10)
4. Loosen the front locking screw ③ on the fluorescence cube
slider. (Fig. 11)
5. Insert the fluorescence cube into the dovetail ④ of the cube
slider and move it to the click position. (Fig. 12)
6. Insert the cube connection cable into the illuminator.
7. Tighten the locking screw ③. (Fig. 11)
Page 13
F ig. 9
F
ig. 9
F ig. 10
F
ig. 10
F F ig. 11
ig. 11
F ig. 12
F
ig. 12