3–8–1 GAS SPRING REPLACEMENT (continued)
CLAMP PLATE
STUD
10. Remove the bellows both from the monitor arm and the main unit. Some force may have to be applied to loosen
the adhesive holding the bellows.
11. Unscrew the four Allen screws to remove the monitor arm from the pipe.
12. Unscrew the screws from the front cover, front plate, front panel, rear cover, and remove them so that the inside of
the main body will be exposed.
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
SCREW
FIXING SCREWS
BOTTOM VIEW – MONITOR ARM
ILLUSTRATION 3–14
PIPE
BELLOWS

RAISED MONITOR ARM

ILLUSTRATION 3–15
3–24
MONITOR ARM
PUSH BRACKET
MONITOR ARM
ALLEN SCREWS
INSTALLATION AND REPLACEMENT