Maximum Allowance Limit For Ecg Leakage Current; Data Sheet For Ecg Leakage Current - GE RTfino/RT3200 Service Manual

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 21
3–7–5 ECG Leakage Current Test (Continued)

Data Sheet for ECG Leakage Current

Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the
table below (TABLE 3–8).
Patient Lead to Ground Leakage Current Test
Patient Lead to Lead Leakage Current Test
Patient Lead Isolation Current Test
Patient Lead Isolation Current Test
Patient Lead to Lead Leakage Current Test
ECG
Tester
Power
Polarity
Patient Lead to Ground Leakage Current Test
Switch
ECG
Power
ON
NORM
ECG
Power
ON
REVERSE
ON
ON
NORM
ON
ON
ON
REVERSE
ON
ON
OFF
NORM
ON
ON
OFF
REVERSE
OFF
ON
OFF
NORM
OFF
OFF
OFF
REVERSE
OFF
OFF
OFF
OFF
OFF
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
TABLE 3–8

MAXIMUM ALLOWANCE LIMIT FOR ECG LEAKAGE CURRENT

and
TABLE 3–9
DATA SHEET FOR ECG LEAKAGE CURRENT
Tester
Neutral
Switch
RL
Tester
Tester
Polarity
Neutral
CLOSED
Switch
Switch
Tester
Tester
Polarity
Neutral
CLOSED
NORM
CLOSED
Switch
Switch
OPEN
REVERSE
CLOSED
NORM
CLOSED
OPEN
NORM
OPEN
REVERSE
CLOSED
CLOSED
REVERSE
OPEN
NORM
OPEN
CLOSED
NORM
CLOSED
REVERSE
OPEN
OPEN
REVERSE
CLOSED
NORM
CLOSED
OPEN
NORM
OPEN
REVERSE
CLOSED
REVERSE
OPEN
NORM
OPEN
REVERSE
OPEN
3–20
MAXIMUM
ALLOWANCE LIMIT
AC POWER
SOURCE
GROUND
OPEN
10 µA
115V
500 µA
220/240V
AC POWER
MAXIMUM
SOURCE
ALLOWANCE LIMIT
115V
220/240V
Tester Lead Selector
RA
LA
LL
Tester Lead Selector
RL
RA
LA
Tester Lead Selector
RL
RA
LA
INSTALLATION AND REPLACEMENT
P9030GX
GROUND
CLOSED
10 µA
10 µA
20 µA
5mA
C
LL
C
LL
C