GE RTfino/RT3200 Service Manual page 90

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 21
SECTION (B) Installation of the French ROM
4.
Remove the BACK PANEL to access the MST board. Refer to illustration 3–35.
4–1. Remove the six rubber caps.
4–2. Unscrew the six screws from the back panel of the RT–x200.
4–3. Remove the back panel from the console.
5.
MST Board
5–1. Remove the MST Board (A3) from the card cage and confirm that the MST Board # is in the list below:
5–2. Remove the English ROM(s) at the location(s) listed in step 5–3.
5–3. Install the German ROM(s) in the following location(s) identified in the Table. Refer to illustration 3–36.
MST Board #
P9514ZZ–00
2113953
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
SCREWS
BACK PANEL REMOVAL
ILLUSTRATION 3–35
CAUTION
Be sure the EPROM pins are straight before installation!
Be very careful to straighten pins if necessary.
Location
English ROM Part #
U9A
P9514NW–00
U11A
P9514NX–00
U9A
2113967–3
U11A
2113968–3
BACK PANEL
SCREWS
German ROM Part #
P9514QV
P9514QW
2113971–3
2113972–3
3–48
INSTALLATION AND REPLACEMENT