10–2 PERIODIC MAINTENANCE PROCEDURE (Continued)
7.
Gently clean video monitor.
Equipment damage possibility. Do not use a hydrocarbon glass cleaner on a monitor with an
anti-glare shield. Prolonged use of such cleaners damages the shield.
7–1. Remove the CRT (VDT) filter from the monitor.
a. Slide the left filter clamp leftwards and pull it out.
b. Slide the right filter clamp rightwards and pull it out.
c. Remove the CRT (VDT) filter.
7–2. Wipe the monitor face and CRT filter using a soft, non-abrasive folded cloth with a glass cleaner.
7–3. Install the CRT (VDT) filter after cleaning the monitor face and CRT (VDT) filter. Set the CRT (VDT) filter on
the monitor, insert the CRT filter clamps ( 1 ), and slide them inwards securely ( 2 ). Refer to ILLUSTRA-
TION 10–1.
Make sure that the filter is securely fixed by the filter clamps. These are important to prevent a hazard
from the filter falling. A "click" sound will be heard when the filter clamp is secured by the lock mecha-
nism.
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
CAUTION

INSTALLING CRT FILTER AND CLAMPS

ILLUSTRATION 10–1
Note
10–4
P9030GX
PERIODIC MAINTENANCE