2–3–5 Facility Power Receptacle
A separate power outlet with a 15 amp circuit breaker for 115 VAC units, or a 10 amp circuit breaker for 220 VAC units,
is recommended. The specific power receptacle used depends on your country's power line standards.
For U.S. applications, use an Underwriters Laboratory (UL) approved Hospital Grade receptacle. For Canadian ap-
plications, the receptacle should be approved by the Canadian Standards Association (CSA). For other countries, the
receptacle should have International Electrotechnical Commission (IEC) approval.
2–4 PHYSICAL SPECIFICATIONS
The RTfino/RT3200 Advantage–II/RT–x200 (excluding accessories) weighs 86 kg (190 lbs). See Illustration 2–2 for
dimensions.
2–4–1 Operating Conditions
This unit is designed to operate within a temperature range of 50 degrees F to 104 degrees F (15.6 degrees C to 40
degrees C), and a relative humidity range of 30 % to 75 %. (Non–condensing).
Patient Comfort
Despite permissible operating temperature and humidity tolerances, we recommend that ambient room temperature
be maintained between 68 and 79 degrees F (20 to 26 degrees C), Humidity should be maintained between 50 % and
70 % for patient comfort during ultrasound scanning.
Electromagnetic Interference (EMI)
In general, the maximum EMI noise Field Strength Intensity should be no more than 100 dB uV/m within the 10 kHz to
100 MHz frequency range, as measured three meters from the this unit. If there is reason for concern, we recommend
that you measure EMI using a spectrum analyzer.
1.
EMI sources should be located and their effects minimized. These noise sources include microwave ovens, cer-
tain types of electronic surgical equipment, and digital equipment such as desk top computers and terminals, gel
warmers, and AM radio broadcast sources.
2.
Keep the unit at least 15 feet (approximately five meters) away from the above noise sources; preferably in anoth-
er room since wood or plaster walls, doors, floors, and ceilings provide some EMI shielding. Special EMI shield-
ing may be required to eliminate noise problems caused by high frequency, high powered radio or video broad-
cast signals.
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
2–6