SECTION (A) Data Recording (continued)
1.
Preparation for Installation
Perform functional checks to verify that the System is functioning properly. Refer to Functional Check Chapter 4
in this Service Manual.
2.
Record or identify all parameters set by the user. Refer to Operator Manual to record or identify.
2–1. Record the Hospital Name.
i Record the Hospital Name, which is displayed in the upper–left corner of the monitor.
2–2. Record Output (Gamma Curve Selection).
i Identify whether the user was using video or polaroid. If the user was using polaroid , measure the Film
Development Time.
Measure the Film Development Time.
i Press the <Control> ,R,P and then the <Enter> key.
ii Press the <Record> key and then measure the Film Development Time and record it.
iii Press the <Control> key.
2–3. Identify the Inverse Display Setting.
i Record the Monitor Display inverse setting (negative or positive).
ii Press the <Record> key and then note the Inverse setting of the printout recorder. If the Monitor
Display changes polarity ( negative or positive) then the printout recorder is set opposite to the monitor.
2–4. Identify the Map Package.
i Read the Operator Manual carefully for this function to avoid confusion.
ii Press the <Control> and N keys and then record whether the Map is A or B.
iii Press the <Control> key.
2–5. Identify the Dotted line.
i Measure the length of something and then record whether the system Dotted Line is ON or OFF.
2–6. Identify the Frame Averaging.
i Watch the image carefully to compare the frame averaging condition before and after step ii is
performed.
ii Press the <Control>, W and <Enter> keys and then record whether the Frame Averaging is ON or
OFF.
iii Press the <Control> key.
2–7. Record the OB Table Data set by the user.
i. Press the <Control> key.
ii Input "G1" followed by the <Enter> key.
iii Record all the OB Table Data set by the user and press the <Enter> key.
iv Repeat steps ii and iii for each of G2,G3,G4 and G5.
v Press the <Control> key.
2–8. Identify the Body Pattern Package in use.
i Press the <
ii Record the Body Pattern which is displayed in the lower–left corner of the monitor.
iii Press the <