2. Power Flatness Performance Wst
The source power level is tested at 201 frequencies across the frequency range
of the analyzer.
Required Equipment and lbols (HP 8719DBOD Only)
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8485A
Adapter 3.5~mm (f) to 3.5~mm (f) . . HP P/N 85052-60012 (part of HP 85052B/D)
Required Equipment and 'Idols (HP 8722D Only)
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 437B, HP 436A, or HP 438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8487A
Adapter 2.4-mm (f) to 2.4nun (f) . . . . . . . .HP P/N 119OOB (part of HP 85056B/D)
1. Zero and calibrate the power meter.
2. Set the calibration factor on the power meter to the average value of the
power sensor between 0.050 and 20 GHz (40 GHz for HP 8722D).
and 92% at 20 GHz (40 GHz for HP 8722D), set the calibration factor to
96%.
Connect the equipment as shown in F'igure 2-18.
3.
NETWORK ANALYZER
H P 8485A
POWER SENSOR
System Verifhtion and
237