Agilent Technologies HP 8719D Service Manual page 408

Network analyzers
Table of Contents

Advertisement

Display Lamp and Assembly
Tools Required
n
n
small slot screwdriver
n
n
Removal
1. Remove the front panel assembly (refer to "Front Panel Assembly" in this
chapter).
2. Remove the three screws (item 1) that attach the display to the front panel.
3. Remove the four screws (item 2), disconnecting the accessories from the
display.
4. Disconnect the cable (item 3) from the Al assembly.
5. Disconnect the display lamp cable (item 4).
6. Lift the display from the front panel and remove the three screws (item 5)
from the outside of the display.
7. Pull the lamp (item 6) out with a curving side motion, as shown.
Replacement
1. Reverse the order of the removal procedure.
14-12
Assembly Replacement and Post-Repair Procedures

Advertisement