0.15 dE3. lb eliminate the calibration factor approximation as the cause of
b. Set the calibration factor on the power meter to the value indicated by
the power sensor.
c. The corrected power level reading should be between the limits shown in
the "Performance l&t Record. B
4. The source relies on the power adjustment for proper performance. Refer to
"Adjustments" for additional information.
5. In case of a repeated failure, after you have made the power adjustments,
refer to "Source Troubleshooting."
and rotate the knob to the frequency in question.