SAVE FAILED. INSUFFICIENT MEMORY
Error Number You cannot store an instrument state in an internal register due
151
to insufficient memory. Increase the available memory by
clearing one or more save/recall registers and pressing [PRESET),
or by storing files to a disk.
SELF TEST #n FAILED
Service Error Internal test #n has failed. Several internal test routines are
Number 112 executed at instrument preset. The analyzer reports the first
failure detected. Refer to the internal tests and the
self-diagnose feature descriptions earlier in this chapter.
SOURCEPOWERTURNED OFF,RESETUNDERPOUERMENU
Information You have exceeded the maximum power level at one of the
Message
inputs and power has been automatically reduced. The
annotation PU indicates that power trip has been activated.
When this occurs, reset the power and then press m !$&jKl!
follows error numbers 57, 58, and 59.
SWEEPMODECHANGEDTO CWTIME SWEEP
Error Number If you select external source auto or manual instrument mode
187
and you do not also select CW mode, the analyzer is
automatically switched to CW.
Service Key Menus and Error Messages
1 O-53