INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE
3.
Select the Device setup tab.
4.
Select the Status tab.
-
An arrow over the picture of the battery indicates the M300 is
on the bedside charger.
-
Bars inside battery indicate strength.
-
Signal strength is displayed in the bar graph.
-
For ECG Electrode status, green indicates the electrode is con-
nected to the patient, red indicates the electrode is not con-
nected to the patient.
ICS MAIN SCREEN BATTERY CHARGING CHECK
The following icons appear in the header bar of the Viewport above
the parameter box & in the header bar of the Bedview next to the pa-
tient category:
M300 on bedside
The bars inside the battery indicate the remaining charge of the M300
battery:
TO RELEARN FOR A M300 or BEDSIDE MONITORED PATIENT
1.
Open the BedView page by selecting the viewport of the pa-
tient on the ICS Main screen.
2.
Select the ECG relearn button on the Bedview menu bar.
3.
For bedside monitored patients you may also select Resp re-
learn to relearn respirations.
Note: It is recommended to INITIATE A MANUAL RELEARN when:
-
Leads are reconnected or electrodes are repositioned
-
Eight hours have passed since the last learning phase
-
The bedside monitor displays a message indicating that it
cannot learn.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
M300 NOT on bedside charger:
charger:
© 2014 Draeger Medical, Inc.
| 25