INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE
TO PAUSE ALL ACTIVE AUDIO ALARM SIGNALS OF AN INDIVIDUAL
PATIENT'S BEDSIDE MONITOR/M300 & ICS*
In the waveform area of the viewport of the bedside moni-
-
tored patient, select the Audio pause button (time length de-
pendent on bedside setting):
In the menu bar of the BedView page of the bedside-monitor
-
patient, select the Bed Audio Pause button
(time length dependent on bedside setting).
Note: Based on the model, software version & settings the follow-
ing may conditions may occur: The ICS, M300 & bedside monitor
may respond differently to the Audio paused state. Default alarm
volumes can be set independently for High, Medium and Low-
priority alarms.
TO SUPPRESS ALL AUDIO & VISUAL ALARM SIGNALS FOR A M300
PATIENT FOR A PRE-CONFIGURED TIME
In the menu bar of the BedView page of the M300 patient, se-
-
lect the Alarm pause button.
-
To cancel before countdown complete select the Alarm pause
button, countdown clock will disappear.
Must be completed at bedside for bedside monitored patients.
-
WHEN THE SPEAKER VOLUME AT THE BEDSIDE MONITOR OR THE
M300 HAS BEEN SET TO ZERO
The following symbol (in yellow) appears in the header bar of
-
the viewport, above the parameter box:
EXTERNAL AUDIO ALARMS
When external audio alarms such as ventilators and incubators
-
(sometimes referred to as MIB) are disabled (set to zero) at the
ICS, the following symbol (in orange) appears in the upper right
corner of the patient's viewport:
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
OR
© 2014 Draeger Medical, Inc.
| 17