Console Full Function Testing
Console Testing to IT0032 Issue 2
6. 23kHz Hand pieces output measurements
The process for the 23kHz straight and angled hand pieces are
identical to the 36kHz. However, the values are different as shown
23kHz Straight hand piece
Q Power
70 Watts Max
Frequency
23kHz Angled hand piece
Q Power
45 Watts Max
Frequency
Max Power
-
Between 22.8kHz and23.2kHz
Max Power
-
Between 22.8kHz and23.2kHz
CUSA EXcel
Res power
30 Watts Min
Res power
20 Watts Min