Console Full Function Testing
Console Testing to IT0032 Issue 2
5.4 Irrigation Measurements
(continued)
Set 3cc/min on the irrigation display in the
STANDBY state
Connect an irrigation tube from an
irrigation source in to a measuring jug
Switch the console to the RUN state
depress and hold on the 'fast flush' pedal
of the footswitch for 60 seconds then
release the pedal and return the console to
the STANDBY state
Irrigation flow rate should be
> 25cc/min.