CUSA EXcel
Console Full Function Testing
Console Testing to IT0032 Issue 2
5.1 Vacuum pump speed adjustment
On the suction pump motor controller
adjust potentiometer R18 (R12 on older
PCB's) For the correct speed with the
console in the RUN state
DO NOT adjust the other potentiometers