Default Menu; Scale Calibration - Rice Lake 150-10-7 Technical Manual

Digital physician scale floor-level
Hide thumbs Also See for 150-10-7:
Table of Contents

Advertisement

150-10-7 Digital Physician Scale
4.5

Default Menu

The default menu is used to return the scale back to its factory settings and is shown below.
PROG
Use the following steps to return the settings back to their factory default.
1. Press
. The display shows a default value of
ENTER
2. To change to YES, use
3. Press
and the display shows DONE.
ENTER
4. Press
to return to top level level menu.
BMI
Selecting YES and pressing
Note
return you to Weigh mode.
4.6

Scale Calibration

Use the following steps to calibrate the scale.
1. Press
and
ENTER
2. The unit cycles through its startup function and continues to display the software version. Continue to hold both keys
until Id appears.
3. Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper
clip, small screwdriver or other similar object to press the setup switch on the back of the indicator (under the cover).
4. Prog displays. Press
5. Calib displays and enter the calibration parameters.
6. Press
and a numeric value is displayed which represents the amount of weight that is used for calibration.
ENTER
Lb will be flashing. To switch between lb and kg, press
right most digit will be flashing.
7. To change the calibration load value, press
8. Use
to move the flashing digit to the left or right.
BMI
9. Once all the digits have been entered, press
10. Make sure the scale platform is clear of weight and press
11. A request to put the chosen load on the platform is displayed by Put xxx.xx.
12. Put the chosen weight on the platform and press
14
Visit our website
CALIB
DEF
NO
YES
Figure 4-4. Default Menu
NO.
and
to adjust the value.
will reset to factory defaults settings without changing the calibration and will
ENTER
simultaneously to power on the scale.
to toggle along the parameter menu.
BMI
or
ENTER
ENTER
www.RiceLake.com
DONE
or
. Once a unit is selected, press
to increment/decrement the flashing digit.
and Clear displays.
again then ===== displays.
ENTER
. ===== displays then Save displays.