Rice Lake 150-10-7 Technical Manual page 16

Digital physician scale floor-level
Hide thumbs Also See for 150-10-7:
Table of Contents

Advertisement

150-10-7 Digital Physician Scale
Parameter
Description
ASTART
Weight process start limit — Maximum (full
capacity)/10; Determine when the weight
algorithm starts (when the "----" is dis-
played), below this value the scale will
show live weight
ARW
Auto Reweigh — Restarts the weight algo-
rithm if the weight changed by more than
this value.
SAL
Semi Auto Live —This value is the interval
between weight displays during the algo-
rithm process
ROUND
Scale Resolution —
Values in kg: 1, 2, 5, 10, 20, 50, 100
Values in lb: 1, 2, 5, 10, 20, 50, 100, 200
DISP
BAUD
Baud rate
ATOL
Algorithm initial tolerance — Maximum
value is 255.
Values above 255 will not let you proceed
and will return to the previous value.
12
Visit our website
Choices
Value (2.0)
Value (4.0 lb)
Value (0.5)
0.2
0.5
1.0
2.0
5.0
10.0
20.0
0.1
0.0
0
0.0000
0.000
0.00
96
48
1152
576
384
288
192
144
Value (10)
Table 4-1. Configuration Mode Menu (Continued)
www.RiceLake.com
Steps
The display toggles between a numeric value and ASTART;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ASTART to ARW; If you want to change the
value, use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. Press BMI key to move to the left.
4. Use the Up/Down arrow keys to increment/decrement numbers.
5. When done, press ENTER key to move to the next parameter (ARW).
The display toggles between a numeric value and ARW;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ARW to SAL; If you want to change the value,
use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. Press BMI key to move to the left.
4. Use the Up/Down arrow keys to increment/decrement numbers.
5. When done, press ENTER key to move to the next parameter (SAL).
The display toggles between a numeric value and SAL;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from SAL to ROUND; If you want to change the
value, use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. When done, press ENTER key to move to the next parameter (ROUND).
The display toggles between a numeric value and ROUND;
The decimal point location is set to the DISP parameter display decimal
point location;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ROUND to DISP; If you want to change the
value, use the following steps;
1. Press ENTER key to change value.
2. Press the Up/Down arrow keys to change the available parameters.
3. When done, press ENTER key to move to the next parameter (DISP).
The display toggles between a numeric value and DISP;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from DISP to BAUD; If you want to change the
value, use the following steps;
1. Press ENTER key to change value.
2. Use the Up/Down arrow keys to change the available parameters.
3. When done, press ENTER key to move to the next parameter (BAUD).
Indicator display illustrates first two digits of baud rate only;
The display toggles between a numeric value and baud;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from BAUD to ATOL;
If you want to change the value, use the following steps
1. Press ENTER key to change value.
2. Use the Up/Down arrow keys to change the available parameters.
3. When done, press ENTER key to move to the next parameter (ATOL).
The display toggles between a numeric value and ATOL;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ATOL to ALEN; If you want to change the
value, use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. Press the BMI key to move to the left.
4. When done, press ENTER key to move to the next parameter (ALEN).

Advertisement

Table of Contents
loading

Table of Contents