Parameter
Description
ALEN
Algorithm initial exponent — Maximum
value 10.
Values above 10, will not let you proceed
and will return to the previous value.
ATOUT
Algorithm maximal exponent — Maximum
value is 15.
Values above 15, will not let you proceed
and will return to the previous value.
TOFF
Auto off timer — Measured in minutes;
0 = always on;
Maximum is 9 minutes;
When using an external power supply, this
parameter is irrelevant
UNITS
Units — Selects the unit of measure; It can
be either Kg/Lb, Kg only or Lb only
OP
Binary options:
OP0 — Live weighing options (0=disable,
1=enable)
OP1 — Communication protocol (0=ESC,
1=maintenance)
OP2 — BMI menu (0=disable, 1=enable)
OP3 — RTC power (0=disable, 1=enable)
OP4 - Semi-Auto-Live — (0=disable,
1=enable)
OP5 - Full calculation — (0=spatial, 1=mul-
tiply by (0=disable, 1=enable
OP6 - Tare - (0=disable, 1=enable
OP7 - Bat type — (0=dry batteries,
1=rechargeable batteries)
OP8 — OIML mode - (0=disable, 1=enable
Press the BMI key to cycle back to the top level menu choices, ie: PROG/CALIB/DEF/DONE
Choices
Value (8)
The display toggles between a numeric value and ALEN;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ALEN to ATOUT; If you want to change the
value, use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. When done, press ENTER key to move to the next parameter (ATOUT).
Value (10)
The display toggles between a numeric value and ATOUT;
If you do not want to change this value, press the BMI key to move to the
next setting; Example: from ATOUT to TOFF; If you want to change the
value, use the following steps;
1. Press ENTER key to change value starting with the right most digit.
2. Use the Up/Down arrow keys to increment/decrement numbers.
3. Press BMI key to move to the left.
4. Use the Up/Down arrow keys to increment/decrement numbers.
5. When done, press ENTER key to move to the next parameter (TOFF).
5
The display toggles between a numeric value and TOFF;
4
If you do not want to change this value, press the BMI key to move to the
3
next setting; Example: from TOFF to UNITS;
2
Press the ENTER key to move to the next parameter; (UNITS); If you want
1
to change the value, use the following steps;
0
1. Press the ENTER key to change values.
9
2. Use the Up/Down arrow keys to change the available parameters.
3. When done, press ENTER key to move to the next parameter (UNITS).
8
7
6
KG/LB
The display toggles between unit of measurements and UNIT;
KG
If you do not want to change this value, press the BMI key to move to the
LB
next setting; Example: from UNITS to OP; If you want to change the value,
use the following steps;
1. Press the ENTER key to change values.
2. Press the Up/Down arrow keys to change the value.
3. When done, press the ENTER key to move to the next parameter (OP).
OP0
The display toggles between a binary option and POO-0;
If you do not want to change this value, press the BMI key to move to the
OP1
OP2
next settings If you want to change the value, use the following step
OP3
1. Press ENTER to change parameters.
2. Use the Up/Down arrow keys to select the display value you want to
change.
3. Press BMI key to move the flashing cursor
4. Press ENTER key to save all of the display parameters. SAVE appears
on the display.
5. Press ENTER key again and DONE appears indicating that you are now
done entering all of the parameters of the scale.
Table 4-1. Configuration Mode Menu (Continued)
© Rice Lake Weighing Systems ● All Rights Reserved
Steps
a. Use the Up/Down arrows to change the value.
b. Press the BMI key to move the flashing cursor.
Configuration
13

Advertisement

Table of Contents
loading

Table of Contents