Hitachi CHROMASTER 5160 Instruction Manual page 90

Table of Contents

Advertisement

1
Program number can be from 1 to 9.
2
When you want to define a new program, enter 1.
As a result, you will delete the previous program.
For modifying the contents of program, enter 0.
3
If pump pressure rises beyond the level set here, the pump
automatically stops.
Set pressure can be from 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Set pressure can be from 0.0 to 60.0 MPa.
5
The set time can be from 0.0 to 600.0.
6
Flow rate valid entries are from 0.00 to 5.000
7
Input the event code. At the set time point, you can stop
the pump, generate an audible signal, markers, timers, etc.
When properly interfaced, these signals can be output to an
external system. No display will show on the screen
unless an event is set. When an event is not set, press only
the ENT key.
8
Repeat steps
9
Program generation will be terminated by pressing ESC
and the monitor screen will reappear.
For confirming the input conditions, check the parameters at each
time point by using
3 - 53
5
7
to
until you complete the program.
keys.