Hitachi CHROMASTER 5160 Instruction Manual page 227

Table of Contents

Advertisement

3. Items to be Prepared by Customer
3.1 Power Supply Arrangement
(f)
Shall be free from excessive dust.
Otherwise, performance degradation might be caused.
(g) Shall be free from a sudden change of power supply
voltage. Otherwise, noise might be caused.
(h) Power supply to motor-driven equipment (such as
stirrer and vibrator) without a noise suppressor shall
not be turned ON/OFF frequently on the same power
supply line as that of this instrument.
(1)
Power supply voltage: 100 to 115 V AC/220 to 240 V AC
Fluctuation shall be ±10% of the rated voltage.
(2)
Frequency: 50 or 60 Hz
Fluctuation shall be ±0.5 Hz of the rated frequency.
(3)
Power Capacity: 450 W
Confirm the necessary power capacity taking actual system
configuration and OA equipment in the installation room into
account and prepare a power supply having an adequate
power capacity.
(4)
Power outlet:
Power supply shall be obtainable near the place for
installing the system inclusive of this detector.
This detector requires a single grounded power outlet (3-
pole type). One or more outlets are needed for the entire
system.
APPENDIX 1 - 3