Hitachi CHROMASTER 5160 Instruction Manual page 873

Table of Contents

Advertisement

(continued)
RACK SET (X2=178.6, Y2=148.5) mm
∗X=178.6 Y=148.5 Z=0.0
Adjust the last vial position in the following way.
(i)
The needle moves to the X2/Y2 position of the
specified rack code. The Z position is relocated at
the uppermost point.
(ii)
Move "∗" to X by the SELECT key and adjust X
finely by moving the needle using arrow keys. Also
using the SELECT key, move "∗" to Y and finely
adjust Y.
(iii) Move "∗" to Z and insert the needle into a vial by the
arrow key.
(iv) Perform fine adjustment in the same procedure as
in 2) and 3).
(v) Press the ENT key to terminate.
NOTE: For fine adjustment in a vial with Z movement,
be sure to perform X/Y adjustments to make
sure the needle can be inserted into the vial
before Z movement into the vial. The Z shift
value in this adjustment will be taken into the
rack parameter. So be sure to perform Z. For
adjustment with the needle inserted, attention
should be paid to the moving distance so that
the needle will not be broken.
MOVING
IDLE WAIT!
Monitor screen reappears.
3 - 29