Hitachi CHROMASTER 5160 Instruction Manual page 67

Table of Contents

Advertisement

STEPS
PROG-NO.
1
1
Input program No. (PROG-No.), repetitive cycles (CYCLE) and conditioning time (COND.
TIME).
STEPS
PROG-NO.
2
1
The conditions for 2nd step are settable here.
Input program No. (PROG-No.), repetitive cycles (CYCLE) and conditioning time (COND.
TIME).
Input the link conditions for the number of steps to be executed. The maximum number of
steps is limited to 9.
ESC
Monitor screen
CYCLE
CONDITION TIME
1
CYCLE
CONDITION TIME
1
The link condition input screen has the following items.
Item
STEPS
Execution step No. cannot be input. No. is counted
sequentially up to 9 max.
PROG-NO.
Input the number of the program to be linked.
Input is valid within a range from 1 to 9.
CYCLE
Input the number of program repetitions. Input is
valid within a range from 1 to 99.
COND. TIME
Input a conditioning time for an execution step.
Input is valid within a range from 0.0 to 600.0 (min).
During the conditioning time, liquid is delivered under
the initial condition (TIME = 0.0) of the execution step.
When the start signal is input after the conditioning
time, program execution starts.
NOTE:
When the same step is repeated (CYCLE value is 2
or larger), conditioning time will be carried out upon
transition to the next cycle after the end of the
previous one.
For analysis by repeating the same step, a column
conditioning time is required. Because conditioning
time affects the reproducibility of retention time, it
should be adequately long. Input a conditioning
time by a conditioning program at the end of the
program specified for the relevant step.
3 - 30